NDJ-1型旋转式粘度计使用说明书NDJ-1 Rotational Viscometer Instruction Manual产品应用:Application本产品广泛应用于测定油脂,油漆,塑料,食品,xx,胶粘剂,化妆品等各种流体的粘性阻力与液体的{jd1}粘度。在使用时必须按照说明书的要求进行安装和测量工作。NDJ-1 Rotational Viscometer is a novel instrument to be used in measuring the viscous resistance and the absolute viscosity of fluid.It is widely applicable to determine the viscosity of various kinds of fluids, such as grease, paint, plastics, food products, pharmaceuticals, bonding agent, etc.1.主要技术规格:   测量范围:10~ 100000 map.s   转速:    6,12,30,60 Spin/minute   测量误差:±5%(牛顿液体)   电源:    220V±10V,50HzMain Technical Specification:   Range:Medium Viscosity 10-100,000mPa.s           Low Viscosity      0.1-100mPa.s           Medium Viscosity   4 Rotors of No.s 1,2,3 and 4.           Low Viscosity      1 Rotor of No.0   Speed of Rotor:6rpm,12rpm,30rpm and 60rpm.   Measurement tolerance: ±5%(Newtonian fluid)   Power Supply:220V±10%V.50Hz2.安装:1) 从包装箱中取出机箱底座与支柱。2) 将三只水平调整脚旋入底座螺丝孔中(图一)3) 主机安装:a. 旋松机箱中螺钉取出主机。b. 旋松手柄固定钮,将电源线嵌入手柄夹头槽内c. 将手柄穿入夹头槽底圆孔中d. 旋紧手柄固定钮夹紧手柄,并使主机基本保持水平e. 拆除指针主杆上的橡皮圈和主机下端的黄色包装套圈。警告!在未拆除上述两项前不可通电开机,否则将损坏机器。f. 接上电源,调整水平调整脚,使水准器中水平泡居于中心圆圈中。Installation:1) Take out the support stand, column from the packing box.2) Screw the column into the hole at the rear of the support stand with the teethed side facing the front of the support stand. Tighten the 3 adjustments underneath the stand support.3) Installation of Main Body:a. Take out of the main body and loose the screw.b. Loose the fixed button and embed the power cord into the groove of the clamp.c. Let the handle get through the hole of groove of the clamp.d. Tighten the fixing screw on lever and keep the main body horizontal.e. Remove the rubber band on the indicator control lever, release the yellow screw at the lower end of the instrument and lastly remove the yellow packing ring. Warning!Do not power on before the 2 operations above, otherwise damage will be caused.f. Power on, make adjustment of the support stand and make sure the leveling bubble is in the center.3. 操作:1) 首先将保护架右旋上仪器下端。2) 准备被测液体,选择直径>or=70mm 的圆柱型容器。3) 按测量要求的精度,准确控制被测液体的温度4) 估计被测液体黏度值的大致范围参考量程表选好转子号,并将转子左旋入连接螺杆(图4).然后旋转升降钮(图5),使仪器慢慢地下降,逐渐浸入测量液体中,直至转子液面标志和液面平为止。(图4)按下指针锁定杆,开启电源开关,放开指针锁定杆,使转子在液体中旋转,一般20~30秒的运转,待指针趋于稳定。按下指针锁定杆,使读数固定下来,再关闭电机,使指针停在读数窗范围内,直接读出数据。当电机关停后,如指针不处于读书窗内,应继续按住指针锁定杆,反复开启和关闭电机,使指针处于读数窗内。如选用6和12转/分时,可直接读取读数,不需锁定指针。5) 读数指针一般在30~90分度之间为佳,否则可变换转子号和转速达到这个要求。6) 当不能预知被测液体粘度值时,应试用由小到大的转子和转动变速钮(图5),由慢到快的变化转速顺序来测定粘度值。Operation:1) Firstly,Spin the protector pole to right just right under the instrument.2) Prepare the fluid to be measured ,put it in a beaker or cylindrical container with its diameter not less than 70mm.3) According to the requested precision of measurement, the temperature of the fluid to be measured should be precisely under control.4) First estimate the approximate range of viscosity of the fluid to be measured, then select the appropriate rotor and speed according to table of range. Screw the selected rotor into the connecting worm(Turn left to install and right to remove).Turn the elevation knob, slowly lower the instrument, the rotor will gradually dip into the fluid surface, then adjust the leveling speed changing knob making the required speed upward to align with the indication point for speed, release the indicator control rod causing the rotor to rotate in the fluid. After several rotation(Generally 20-30seconds) until the indicator tends to be stable(or make readings at specified time).Press indicator control rod(Caution: 1. Do not use excessive force. 2.When the speed is quite slow, it is not necessary to use control rod. The readings can be directly made), causing the reading to settle, switch off the motor, the indicator will stay at the readout window and then reading can be taken. After the motor has switched off but the indicator does not stay at the readout window, press the indicator control rod continuously and repeatedly switch off the motor. After several practices the user will skillfully manage the technique of keeping the indicator too stay in the window for reading.5) When the value indicated by the indicator is too high or too low, change the rotor or speed so the reading is preferable range of 30-90 graduation.6) When the approximate viscosity of the fluid to be measured is difficult to estimate, suppose a higher viscosity first, then try the rotor of lower No. to higher No. and lower speed to higher speed. The principle is that for the fluid of high viscosity select small rotator and low speed. For the fluid of low viscosity select large rotor and high speed.4. 注意事项:1) 本仪器使用于常温环境下(10-30?C).2) 测定时必须装上保护架,否则测定粘度将因容器的直径变化而变化。3) 转子在安装和拆除时应将转子连接螺杆向上抬起,避免转轴轴尖碰损,浸入液体时避免气泡粘附于转子上。4) 仪器不得测放或倒放,特别是在装上转子时。5) 装上转子后,不要在未浸入液体的情况下,开机旋转。6) 每次使用完毕小心拆下转子,彻底清洗,用吹风机吹干,垂直安放在转子架上。7) 搬运仪器时应装上黄色包装套圈和橡皮圈锁定指针锁定杆。8) 直接测定被测液体的,不要以恒温槽的温度代替被测液体的温度。9) 在测试中非牛顿液体的表现粘度往往因切变速率和测定时间长短等因素的不同而变化,属正常情况,非仪器故障。10) 严格按照使用要求操作使用。Points for Attention:This instrument should be used in regular temperature(Normally 10-30 ?C).The protector should be applied in the course of measurement, otherwise the viscosity will vary with the diameter of the container.Care should be taken in installing and removing the rotor. To install and remove the rotor the connecting worm should be slightly lifted to avoid he shaft tip damaged.When the rotor is installed, the instrument should not be placed on its side or upside down.After installation of rotor do not power on in the case that the rotor is not dipping into the fluid.After each operation, clean the rotor immediately (don't clean the rotor when it is on the instrument). After cleaning, put the rotor on the rotor rack properly.When moving or transporting the instrument, use a rubber band to wrap the indicator control rod and keep it tightening, sleeve it into the yellow packing ring and lift the connecting worm.Precisely control the temperature of the fluid to be measured.Most of the high viscosity fluid, such as suspension, emulsion, high polymer, etc., are usually "Non-Newtonian fluid", their variation of viscosity is according to the shear rate and change of time. Therefore measurement under different rotor, speed and time may give non-uniform results. This is normal, and it is no the fault of the instrument.Strictly adhere to the operation instructions for measurement.

用 途
    NDJ-1型旋转式粘度计是用于测量液体的粘性阻力与液体的{jd1}粘度的新型仪
    器。广泛适用于测定油脂、油漆、塑料、食品、xx、胶粘剂等各种流体的粘度。
    主要技术规格
    测定范围:中粘度    lO-100000mPa·s
    低粘度    O.1-lOOmPa·s
    中粘度    1、2、3、4号四种转子
    低粘度    0号转子
    转子速度:6转/分 12转/分 30转/分 60转/分
    测量误差: +5%(牛顿液体)
    电    源:220V 50Hz
    外形尺寸:300mm x 300mm x 450mm
    净 重:1.5kg(不包括支架)
结构原理

1.如图所示,同步电机以稳定的速度旋转,连接刻度圆盘,再通过游丝和转轴带动转子旋转。如果转子未受到液体的阻力,则游丝、指针与刻皮圆盘同速旋转,指针在刻度盘上指出的读数为“0”。反之,如果转子受到液体的粘滞阻力,则游丝产生扭矩,与粘滞阻力抗衡{zh1}达到平衡,这时与游丝连接的指针在刻度圆盘上指示一定的读数(即游丝的扭转角)。将读数乘上特定的系数即得到液体的粘度(mpa·s)
2.利用齿轮系统及离合器进行变速,由专用旋转钮操作,分四档转速,根据测定需要选择。
3.本仪器有1至4号四种转子,可根据被测液体粘度的高低随同转速配合选用。(0号转子需另购)
4.仪器装有指针固定控制机构,为xx读数用,当转速较快时(30/分,60转/分)无法在旋转时进行读数,这时可按下指针控制杆,使指针固定下来,便于读数。
5.保护架是为稳定测量和保护转子用。使用保护架进行测定能取得较稳定的测量结果。
6.仪器可手提使用,配有固定支架及升降机构,—般在实验室中进行小量和定温测定时应固定。

安 装
1.从包装箱中取出存放箱、支架和支柱等。
2.将支柱旋人支架后部之螺孔中,并将支柱上的齿形面面向支架正前方,用扳头将六角螺母拧紧,防止支柱转动;
3.转动升降夹头旋钮,检查升降夹头的灵活性和自锁性,发现过松和过紧现象可用内六角螺钉扳头调正夹头紧松螺钉,使能上下升降,偏紧为宜,以防装上粘度计后产生自动坠落情况。
4.打开存放箱,取出粘度计,并取下套在粘度计上泡沫盖,把粘度计装在升降夹头上,用手柄固定螺钉拧紧,接上电源,拿下指针控制杆上的橡皮筋,旋松仪器下端螺钉,取下黄色包装套圈。  
5.调节水平调正螺钉,·保持仪器水平。
操作使用
1.准备被测液体,置于直径不小于70nm的烧杯或直筒形容器中,准确地控制被测液体温度。
2.将保护架装在仪器上(向右旋入装上,向左旋出卸下)。
3.将选配好的转子旋入连接螺杆(向左旋人装上,向右旋出卸下)。旋转升降钮,使仪器缓慢地下降,转子逐渐浸入被测液体中,直至转子液面标志和液面平为止,调正仪器水平,开启电机开关,转动变速旋钮,使所需转速数向上,对准速度指示点,使转子在液体中旋转,经过多次旋转(一般20—30秒)待指针趋于稳定(或按规定时间进行读数)。按下指针控制杆(注意:
1.不得用力过猛。
2.转速慢时可不利用控制杆,直接读数)使读数固定下来,再关闭电机,使指针停在读数窗内,读取读数。当电机关停后如指针不处于读数窗内时,可继续按住指针控制杆,反复开启和关闭电机,经几次练习即能熟练掌握,使指针停于读数窗内,读取读数。
4.当指针所指的数值过高或过低时,可变换转子和转速,务使读数约在30~90格之间为佳。
5、使用0号转子和低粘度测试附件可按下列步骤操作。
    a、将0号转子装在连接螺杆上(向左旋转装上)。
    b、将固定套筒套入仪器底部圆筒上,并用套筒固定螺钉拧紧。
    c、将外试筒(有底)内注入20—25ml被测体。
    d、将外试筒套人固定套筒并用试筒固定螺钉予以拧紧,旋紧时必须注意试筒
    固定螺钉之锥端以旋人外试筒上端之三角形槽内(可以侧面之圆孔中观察试筒三角槽
    是否位于圆孔中心)。控制好被测液体温度后即可进行测试。
    e、配用无底外试筒时,方法基本同上(除c),当外试筒和转子浸入液体时以
    固定套筒上之红点作为液面线。
    6.量程,系数及转子、转速的选择:
    a、先大约估计被测液体的粘度范围,然后根据量程表选择适当的转子和转速。
    如测定约3000毫:咱·秒左右的液体时可选用下列配合:
    .2号转子……6转/分
    或3号转子……30转/分
    b、当估计不出被测液体的大致粘度时,应假定为较高的粘度,试用由小到大
    的转子和由慢到快的转速。原则是高粘度的液体选用小的转子、慢的速度;低粘度的
    液体选用大的转子和快的转速。
    c、系数:测定时指针在刻度盘上指示的读数必须乘上系数表上的特定系数才为
    测得的{jd1}粘度(mPa·s)。

即    η=K·α
    式中:η:{jd1}粘度
    K=系数
    α=指针所指读数(偏转角度)
    d、频率误差的修正:当使用电源频率
    不准时,可按下公式修正。
    名义频率
    实际粘度=指示粘度·(名义频率/实际频率)

注意事项
    1.本仪器适于常温环境下使用。
    2.仪器必须在指定频率和电压允差范围内测定,否则会影响测量精度。
    3.尽可能利用支架固定仪器测定。如手持操作应保持仪器稳定和水平。
    4.装卸转子时应小心操作,装拆时应将连接螺杆微微抬起进行操作,不要用力过
大,不要使转子横向受力,以免转子弯曲。
    5.装上转子后不得将仪器侧放或倒放。
    6.一定要在电机运转时变换转速。
    7.连接螺杆和转子连接端面及螺纹处应保持清洁,否则将影响转子的正确连接及
转动时的稳定性。
    8.仪器升降时应用手托住仪器,防止仪器自重坠落。  
    9.每次使用完毕应及时清洗转子(不得在仪器上进行转子清洗)清洁后要妥善安
放于转子架中。
    10。装上0号转子后不得在无液体的情况下“旋转”以免损坏轴尖。
11.使用0号转子时不用保护架。
    12.不得随意拆动调整仪器零件,不要自行加注润滑油。
    13.仪器搬动和运输时应用橡皮筋将指针控制杆圈柱,并套入黄色包装套圈托起
    连接螺杆,然后用螺钉拧紧。
    14、悬浊液、乳浊液、高聚物及其他高粘度液体中很多都是“非牛顿液体”其表
    现粘度随切变速度和时间变化而变化,故在不同的转子、转速、和时间下测定,其结果不一致是属正常情况,并非仪器不准(一般非牛顿液体的测定应规定转子、转速和
    时间)。
    15。做到下列各点能测得较xx的数值:
    a、xx地控制被测液体的温度。
    b、将转子以足够长的时间浸于被测液体同时进行恒温,使其能和被测液体温
    度一致。
    c、保证液体的均匀性。
    d、测定时尽可能将转子置于容器中心。
    e、防止转子浸入液体时有气泡附粘于转子下面。
    f、变换转子或转速使刻度圆盘上的读数偏高些。
    z、使用保护架进行测定。
    h、保证转子的清洁。  
    i、低于15毫帕·秒的液体选用0号转子。
    j、电源频率不准时按修正公式修正。
    k、严格按照使用操作规则进行操作。